Lindberg Tube Furnace 55035 Manual

The Lindberg Tube Furnace 55035 is a high-temperature laboratory furnace designed for various thermal processing applications. This single-zone tube furnace features a maximum operating temperature of 1200°C, digital temperature controller, ceramic fiber insulation, and quartz tube work chamber. Below are key sections for safety, features, installation, operation, maintenance, and troubleshooting.

1. Features at a Glance 2. Installation 3. First-Time Setup 4. Specifications 5. Controls and Settings 6. Connecting Accessories 7. Operating Procedures 8. Safety Features 9. Maintenance & Cleaning 10. Regulatory Information 11. Troubleshooting

Features at a Glance

Key components: High-temperature heating elements, Digital PID controller, Quartz tube chamber, Ceramic fiber insulation, Over-temperature protection, Cooling fan.

FeatureDescription
Heating elementsKanthal wire heating elements for high-temperature operation
Digital controllerPID temperature control with programmable ramps
Quartz tubeHigh-purity quartz work chamber (OD 50mm)
InsulationCeramic fiber insulation for energy efficiency
Power input120V AC, 15A standard outlet
Temperature rangeAmbient to 1200°C maximum
Cooling systemBuilt-in cooling fan for rapid cooldown
Safety featuresOver-temperature protection, door interlock
DisplayLED digital display for temperature and status
Control buttonsSET, UP, DOWN, ENTER for programming

Installation

Unpack and inspect contents. Place on stable, level surface.

  1. Location: Install in well-ventilated area with adequate clearance (minimum 6 inches from walls).
  2. Power connection: Connect to dedicated 120V, 15A circuit with proper grounding.
  3. Tube installation: Carefully insert quartz tube, ensuring proper seating in end caps.
  4. Gas connections: Connect gas inlet/outlet tubes if required for process atmosphere.
  5. Verify installation: Ensure all connections are secure before powering on.

WARNING! Furnace operates at extremely high temperatures; maintain safe distance during operation. Ensure proper ventilation.

First-Time Setup

Power on furnace; follow initial setup procedure.

  1. Initial heating: Run furnace empty at 500°C for 2 hours to burn off manufacturing residues.
  2. Controller setup: Set temperature units (°C/°F) and control parameters.
  3. Program setup: Configure ramp rates and hold times for typical processes.
  4. Safety check: Verify over-temperature protection is functioning properly.
  5. Calibration: Check temperature accuracy with external thermocouple if required.

CAUTION! Wear appropriate PPE (heat-resistant gloves, safety glasses) when operating furnace.

Specifications

Controls and Settings

Access control functions via front panel buttons and digital display.

Front Panel: POWER switch, SET button, UP/DOWN arrows, ENTER button, digital display.

Temperature Control: Set point adjustment, ramp rate programming, hold time setting. Safety: Over-temperature protection setting, door interlock status. Display: Current temperature, set point, program status, error codes. Programming: Multi-segment programs with ramp and hold functions.

Connecting Accessories

Gas flow systems: Connect to gas inlet/outlet ports. Vacuum systems: Connect to vacuum pump via appropriate fittings. Temperature monitoring: External thermocouple connections. Data logging: RS-232 or USB output for process recording.

  1. Turn off furnace before connecting accessories.
  2. Use appropriate high-temperature fittings and seals.
  3. Pressure test gas/vacuum connections before heating.
  4. Ensure all connections are rated for maximum operating temperature.

Tip: Always use high-temperature rated tubing and fittings for gas connections.

Operating Procedures

Standard operation for thermal processing applications.

  1. Loading: Carefully place samples in quartz boat and insert into tube center.
  2. Gas flow: Establish appropriate gas flow rate if using process atmosphere.
  3. Programming: Set desired temperature profile using controller.
  4. Heating: Start program and monitor temperature ramp.
  5. Hold: Maintain at target temperature for required duration.
  6. Cooling: Allow furnace to cool naturally or with forced cooling.
  7. Unloading: Remove samples only when temperature is below 100°C.

WARNING! Never touch hot surfaces. Use proper handling tools and PPE.

Safety Features

Over-temperature Protection: Automatic shutdown if temperature exceeds set limit. Door Interlock: Prevents operation when access door is open. Ground Fault Protection: Electrical safety circuit. Cooling Fan: Prevents overheating of exterior surfaces.

Maintenance & Cleaning

Regular maintenance ensures optimal performance and longevity.

  1. Monthly: Inspect heating elements and thermocouple for damage.
  2. Quarterly: Clean quartz tube with appropriate solvents.
  3. Annually: Check electrical connections and calibrate temperature controller.
  4. As needed: Replace worn gaskets and seals.

CAUTION! Always disconnect power before performing maintenance. Allow furnace to cool completely.

Regulatory Information

UL Listed: Complies with UL safety standards. CE Marked: Meets European Union safety requirements. FCC Part 15: Class A digital device; may cause interference in residential areas. Copyright © 2025 Lindberg; trademarks of SPX Corporation. For technical support, visit www.lindbergfurnaces.com or contact 1-800-873-5658.

Troubleshooting

SymptomPossible CauseCorrective Action
Furnace won't heatPower supply/controllerCheck power connection; verify controller settings; check circuit breaker.
Temperature unstableThermocouple/controllerCheck thermocouple connection; perform PID auto-tune; replace thermocouple if damaged.
Error code displayedVarious faultsRefer to error code table in appendix; reset controller; check safety interlocks.
Quartz tube crackedThermal shock/ageReplace quartz tube; avoid rapid temperature changes; inspect for stress points.
Poor temperature uniformityHeating elements/insulationCheck element continuity; inspect insulation; verify proper tube positioning.

Reset: Power cycle furnace; controller reset procedure if available.

Service: 1-800-873-5658; www.lindbergfurnaces.com