Thermo Scientific Thermolyne Furnace Manual

The Thermo Scientific Thermolyne Furnace is a high-performance laboratory furnace designed for various heating applications including ashing, heat treating, annealing, and materials testing. Featuring precise temperature control, durable construction, and user-friendly operation, it is suitable for research, industrial, and educational environments. Below are key sections covering safety, features, installation, operation, maintenance, and troubleshooting.

1. Features at a Glance 2. Installation 3. First-Time Setup 4. Specifications 5. Controls and Settings 6. Operating Procedures 7. Maintenance 8. Safety Features 9. Care & Cleaning 10. Regulatory Information 11. Troubleshooting

Features at a Glance

Key components: High-temperature chamber, Digital temperature controller, Heating elements, Thermal insulation, Door safety switch, Power cord.

FeatureDescription
Temperature RangeUp to 1200°C (dependent on model)
Digital ControllerPID control with programmable settings
Chamber DimensionsVaries by model (e.g., 6x6x6 inches)
Heating ElementsKanthal or silicon carbide elements
InsulationCeramic fiber for efficient heat retention
Safety FeaturesOver-temperature protection, door interlock
Power Requirements120V/240V AC, 50/60Hz
DisplayLED or LCD digital readout
ProgrammabilityMulti-segment temperature profiles
ConstructionStainless steel exterior

Installation

Unpack and inspect contents. Place on stable, heat-resistant surface.

  1. Location: Install in well-ventilated area away from flammable materials.
  2. Power Connection: Connect to appropriate power supply (120V or 240V).
  3. Grounding: Ensure proper electrical grounding.
  4. Clearance: Maintain minimum 6-inch clearance on all sides.

WARNING! Furnace reaches extremely high temperatures; install away from combustible materials. Use heat-resistant gloves when handling.

First-Time Setup

Power on furnace; follow initial setup procedure.

  1. Initial Burn-off: Run empty at 500°C for 2 hours to remove manufacturing residues.
  2. Temperature Calibration: Verify temperature accuracy using calibrated thermometer.
  3. Program Setup: Configure default temperature programs if applicable.
  4. Safety Check: Test door interlock and over-temperature protection.

CAUTION! Always wear appropriate personal protective equipment (PPE) when operating furnace.

Specifications

Controls and Settings

Access control panel for temperature programming and monitoring.

Control Panel: POWER, TEMPERATURE SET, PROGRAM, ENTER, UP/DOWN arrows.

Temperature Settings: Set point, ramp rate, hold time. Safety: Over-temperature protection, door safety switch. Programs: Multi-segment profiles with ramp and soak functions. Display: Current temperature, set point, program status.

Operating Procedures

Standard operating procedure for furnace use.

  1. Preparation: Load samples using crucibles or boats.
  2. Programming: Set desired temperature profile.
  3. Start: Close door securely and start program.
  4. Monitoring: Observe temperature display during operation.
  5. Completion: Allow furnace to cool before opening door.
  6. Unloading: Use tongs and heat-resistant gloves.

Tip: Always pre-heat empty furnace to desired temperature before loading samples.

Maintenance

Regular maintenance ensures optimal performance and longevity.

  1. Monthly: Inspect heating elements for damage.
  2. Quarterly: Clean chamber with soft brush or vacuum.
  3. Annually: Calibrate temperature controller.
  4. As needed: Replace damaged insulation or elements.

WARNING! Always disconnect power and allow furnace to cool completely before performing maintenance.

Safety Features

Over-temperature protection: Automatic shut-off if temperature exceeds set point. Door safety interlock: Prevents operation when door is open. Ground fault protection: Electrical safety circuit. Thermal insulation: Reduces exterior surface temperature.

Care & Cleaning

Disconnect power and allow to cool completely before cleaning. Use soft brush or vacuum for chamber cleaning. Wipe exterior with damp cloth.

CAUTION! Never use water or chemical cleaners on hot surfaces. Avoid abrasive materials that could damage insulation.

Regulatory Information

UL/CSA certified for electrical safety. CE marked for European compliance. Manufactured in accordance with ISO 9001 quality standards. Copyright © Thermo Fisher Scientific. For technical support, visit www.thermofisher.com or contact 1-800-522-2255.

Troubleshooting

SymptomPossible CauseCorrective Action
Furnace not heatingPower supply/elementsCheck power connection; inspect heating elements; verify controller settings.
Temperature inaccurateCalibration/thermocoupleRecalibrate controller; check thermocouple connection.
Door won't lockSafety interlockCheck door alignment; inspect interlock switch.
Error code displayedController faultRefer to error code in manual; reset controller.
Excessive heat lossInsulation damageInspect chamber insulation; replace if damaged.

Reset: Power cycle furnace; reset controller to factory defaults if needed.

Service: 1-800-522-2255; www.thermofisher.com