The Thermolyne NCAT Asphalt Content Furnace is a specialized laboratory instrument designed for determining asphalt content in hot mix asphalt (HMA) samples using the ignition method. This furnace features digital temperature control, automatic test sequencing, built-in calibration, safety interlocks, and compliance with AASHTO T 308 and ASTM D 6307 standards. Below are key sections for safety, operation, calibration, maintenance, and troubleshooting.
Key components: Digital temperature controller, ceramic combustion chamber, sample basket assembly, exhaust system, safety door interlock, calibration software.
| Feature | Description |
|---|---|
| Digital Controller | Microprocessor-based with LCD display |
| Temperature Range | Up to 600°C (1112°F) |
| Combustion Chamber | Ceramic fiber insulation |
| Sample Capacity | Up to 4000g asphalt mixture |
| Exhaust System | Built-in fume extraction |
| Safety Interlock | Door safety switch |
| Calibration Memory | Stores multiple calibration factors |
| Test Modes | Standard test, calibration, manual |
| Power Requirements | 240V AC, 50/60Hz |
| Compliance | AASHTO T 308, ASTM D 6307 |
Unpack and inspect all components. Ensure proper ventilation and electrical supply.
WARNING! Furnace operates at high temperatures; ensure proper clearance from combustible materials. Use heat-resistant gloves when handling hot components.
Power on furnace; perform initial calibration and safety checks.
CAUTION! Always wear appropriate PPE: heat-resistant gloves, safety glasses, and lab coat during operation.
Front Panel: POWER switch, DOOR OPEN/CLOSE, EMERGENCY STOP, temperature display, function keys.
Controller Functions: TEMP SET, TIME SET, CALIBRATION, TEST START/STOP, MANUAL MODE.
Operation Modes: STANDARD TEST - automatic ignition test; CALIBRATION - establish correction factors; MANUAL - user-controlled temperature and time. Safety Features: Door interlock prevents operation when open; over-temperature protection; emergency stop button.
Proper sample preparation is critical for accurate results.
Tip: Samples should be at room temperature before testing to prevent thermal shock.
Standard test procedure for asphalt content determination.
WARNING! Do not open door during test. Chamber temperatures exceed 500°C and can cause severe burns.
Regular calibration ensures accurate results. Perform when changing aggregate source or every 6 months.
Calibration factors are stored in non-volatile memory and applied automatically to test results.
Regular maintenance ensures optimal performance and longevity.
Daily: Wipe exterior with damp cloth; check exhaust airflow. Weekly: Clean combustion chamber; inspect heating elements. Monthly: Check door seal; verify calibration; inspect electrical connections.
CAUTION! Always disconnect power before performing maintenance. Allow furnace to cool completely before cleaning.
FCC Part 15: Class A digital device; may cause interference in residential areas. UL listed; CSA certified. Always wear appropriate PPE. Ensure proper ventilation to prevent fume accumulation. Keep flammable materials away from furnace. Emergency procedures: Use EMERGENCY STOP button; have Class C fire extinguisher available.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No power | Circuit breaker, power cord | Check circuit breaker; verify power cord connection; ensure proper voltage. |
| Temperature fluctuation | Thermocouple, heating elements | Check thermocouple connection; inspect heating elements for damage. |
| Door won't close | Interlock, obstruction | Check for obstructions; verify interlock switch operation. |
| Inconsistent results | Calibration, sample prep | Recalibrate furnace; verify sample preparation procedure. |
| Excessive smoke | Ventilation, sample size | Check exhaust system; reduce sample size; ensure proper ventilation. |
Reset: Power cycle unit; perform system reset from maintenance menu if available.
Service: Contact Thermolyne Technical Support: 1-800-553-0039; www.thermolyne.com